Z-0885-2025: Welch Allyn, Inc., Life2000 Ventilation System, REF BT-20-0002A

Recall record Z-0885-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0885-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jan 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.

Source facts

Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/20/2024 via USPS first class mail. The notice explained the problem with the device, hazard involved, and requested the following actions be taken: 1. Always have an alternate means of ventilation or oxygen therapy available. 2. Patients may continue using the Life2000 ventilator if it is functioning correctly. If the ventilator becomes inoperable, place the patient on an alternate means of ventilation, if necessary, and contact Baxter Advanced Respiratory, Clinical Support team at 800-397-9071. 3. In the interim if patients experience this failure and have an additional charger available, connect it to the ventilator and verify that the ventilator is charging. 4. If you received this communication directly from Baxter, please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any remaining inventory. Log in to the portal using the account number listed in the enclosed reply form instruction sheet. 6. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to the Director of Respiratory Therapy, Patient Safety, and any other departments within your institution who use the affected product. 7. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal. Durable Medical Equipment providers are required to notify patients of this Urgent Medical Device Correction by forwarding the attached home patient letter. For general questions regarding this Urgent Medical Device Correction, please contact the Baxter Advanced Respiratory team at 800
Initiated
2024-12-20
Root cause
Process control
Distribution
US, Nationwide Distribution
Recall number
Z-0885-2025
Recall status
Open, Classified
CDRH record ID
211338
FDA classified
2025-01-16
Recall event number
95842

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗