Skip to content

Public FDA record

Z-0886-04: LIFEPAK 500 automated external defibrillator

Recall record Z-0886-04 for Medtronic Physio Control Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0886-04
Company in FDA record
Medtronic Physio Control Corp
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for batteries to malfunction due to battery cell rupture.

Source facts

Action
On 4/2/04 the firm initiated telephone contacts to customers to determine if the customer potentially has suspect battery. If so, the customer will be faxed a letter describing how to remove the battery to check for the code date. If customer confirms they have suspect battery, a new battery will be shipped to customer.
Initiated
2004-04-02
Root cause
Other
Terminated
2005-01-06
Distribution
The firm distributed devices nationwide to various hospitals, medical centers, and businesses. The firm also distributed to foreign customers in Australia, Argentina, Brazil, Canada, China, Greece, Hong Kong, Israel, India, Jordan, Mexico, Netherlands, New Zealand, Oman, Philippines, Saudi Arabia, United Kingdom.
Recall number
Z-0886-04
Recall status
Terminated
CDRH record ID
32418
FDA classified
2004-07-20
Recall event number
28727

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗