Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.
Source facts
Action
Firm sent letters on 03/30/2007 dated 03/2007 via certified return receipt to consignees with affected LIFEPAK 20 devices. The firm advised consignees to use the Operator's Checklist in the Operating Instructions to perform daily user tests and function checks while operating the defibrillator on battery power. In addition consignees are advised to replace their battery every 1 to 2 years.
Initiated
2007-03-30
Root cause
Other
Terminated
2011-05-04
Distribution
Distribution is worldwide. The device is designed to be used in healthcare provider facilities such as hospitals and clinic settings.
Recall number
Z-0886-2007
Recall status
Terminated
CDRH record ID
51329
FDA classified
2007-07-04
Recall event number
37692