Z-0886-2015: ACCU-CHEK Tender I 17/60 10 pieces Product Usage: Intended for the infusion and/or injection of fluids i

Recall record Z-0886-2015 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0886-2015
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA classification date
Dec 30, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify

Source facts

Action
Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
Initiated
2014-11-20
Root cause
Device Design
Terminated
2018-01-10
Distribution
US Nationwide Distribution
Recall number
Z-0886-2015
Recall status
Terminated
CDRH record ID
131589
FDA classified
2014-12-30
Recall event number
69821

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗