Open original FDA record ↗Recall record
- FDA identifier
- Z-0886-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Dec 5, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
Source facts
- Action
- Medline Industries issued an URGENT MEDICAL DEVICE RECALL to its consignee on 10/27/2025 via email. The notice explained the issue, potential risk, and requested the affected product be quarantined. Once the firm receives the response form, Medline will provide over-labels which are to be placed on kits that contain the recalled component and will state the affected component shall be removed and discarded from further use. Distributors and those further distributing the affected product are directed to notify those to whom the product was transferred or distributed of the recall. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Initiated
- 2025-10-27
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the state of CA.
- Recall number
- Z-0886-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217344
- FDA classified
- 2025-12-05
- Recall event number
- 97968