Z-0887-05: LifeScan brand OneTouch Ultra Test Strip used in LifeScan Blood Glucose Meter; Part Number 020-244-07; Di

Recall record Z-0887-05 for Lifescan Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0887-05
Company in FDA record
Lifescan Inc
FDA classification date
Jun 14, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A Control Solution range for the OneTouch Ultra Test Strip of 111-150 mg/dL was incorrectly labeled as 97-131 mg/dL.

Source facts

Action
On 11/17/03 the firm initiated the recall and their notification was via letters informing its customers of the removal.
Initiated
2003-11-17
Root cause
Other
Terminated
2005-09-20
Distribution
Product was released for distribution to 30 consignees nationwide. The firm provided the following distribution information: 6 wholesaler dealers, 20 distributors, 3 retailers and one (1) consumers/users involved in the recall. Note: This customer received a shipment of 2 Test Strip boxes (2 boxes x 50 strips/box = 100 test strips) as 'free goods' and was most likely for an LifeScan employee in the field. The lot was released from LifeScan's facility in Puerto Rico on May 1, 2003 and received at LifeScan's warehouse in Milpitas, CA on May 16, 2003. From the Milpitas warehouse, this lot was transferred to LifeScan's warehouse in Mechanicsburg, PA where it was received on June 16, 2003. The affected lot was distributed out of the Mechanicsburg, PA warehouse only. The recall product was distributed in the U.S. only, to the following states: AL, DE, FL, IL, IN, MA, MI, MN, MO, NC, NH, NJ, OH, PA, TN, TX, VA and WV. (A second lot was also mislabeled with the incorrect Control Solution range. This lot was manufactured outside of the US and distributed exclusively in Italy.) The total number of government sales/distribution centers that received the recall product = 0.
Recall number
Z-0887-05
Recall status
Terminated
CDRH record ID
39361
FDA classified
2005-06-14
Recall event number
32200

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗