Skip to content

Public FDA record

Z-0887-2022: Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.

Recall record Z-0887-2022 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0887-2022
Company in FDA record
Maquet Cardiovascular, LLC
FDA record creation date
Apr 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.

Source facts

Action
On February 14, 2022, a customer notification letter was issued via FedEx. Customers are instructed to quarantine and return all affected product and notify all users within the hospital/facility. Please note affected lot numbers appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number, which is etched onto all 3 device components on side of handle and on underside above each pin. Customers are to complete and return the response form by email or fax. Distributors are to forward this document to customers for appropriate action. If you have any questions, please contact your Maquet Cardiovascular, LLC /Getinge representative or call the Getinge Customer Service at (888) 880-2874, Monday through Friday, between the hours of 6:00a.m. and 5:00 p.m. (Pacific Standard Time)
Initiated
2022-02-09
Root cause
Mixed-up of materials/components
Distribution
Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.
Recall number
Z-0887-2022
Recall status
Open, Classified
CDRH record ID
192340
FDA record created
2022-04-12
Recall event number
89744

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗