Z-0888-2017: Medi Choice Arm Slings ASL 2103 Small 7.5 x 12.5 Inch 1/EA, 24 ENCA ASL 2104 Medium 8 x 15 Inch 1/EA, 24

Recall record Z-0888-2017 for Owens & Minor Distribution, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0888-2017
Company in FDA record
Owens & Minor Distribution, Inc.
FDA record creation date
Dec 22, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Owens & Minor is the initial importer and private label distributor of Medi Choice Arm Slings (Product). The Product is manufactured in China by Tops Medical Articles Nantong Inc. Owens & Minor first learned of the product issue that prompted the product removal action via customer complaints received in September 2016. Two customer complaints received on September 1"1, 2016 reported surface contamination on the arm slings. Two additional complaints reported on September 19, 2016 and October 7, 2016 reported the same issue. Upon receipt of the complaints on September 1, 2016, Owens & Minor immediately initiated a CAPA investigation for root cause analysis and determination of corrective/preventive actions. A quality inspection of inventory confirmed that the surface contamination was confined to lots manufactured in May, June, and July of 2016. The surface contamination for these lots was found to be frequent. Subsequent August product lots were inspected and no surface contamination was found. Limited inventory from lots prior to May 2016 were available and inspected as well. No surface contamination was found on these prior lots. Owens & Minor has quarantined the affected product lots. The product is on hold at all Owens & Minor Distribution Centers (DC). Samples of contaminated product have been sent to the manufacturer for evaluation and to a certified testing lab for identification of contaminate. These results are pending at this time. Based on the frequency of the product defect as well as evaluation of potential adverse health event risk, Owens & Minor initiated a voluntary recall for the above stated affected product lots.

Source facts

Action
The firm, Owens&Minor, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 11/9/2016 to its consignees/customers including a Recall Response Form. The letter describe the product, problem and actions to be taken. The customers were instructed to examine your inventory and identify all affected product; immediately stop using all affected product; quarantine all affected product per your facility's Standard Operating Procedures; if you have further distributed this product, identify patients/customers who received product and notify them at once; and complete and return Recall Response Form via email: www.owens-minor.com, fax to 804-723-7100, or mail to the O&M Distribution Center (DC) Attn: Recall Officer (even if you do not have any affected product). Note: upon receipt of the Recall Response Form, the DC will contact you with the appropriate RGA and instructions for product return. If you have any questions concerning credit and/or replacement of affected product or this recall, please call 804-723-7000 or contact your O&M Sales Representative or Customer Service Representative at the local O&M Distribution Center.
Initiated
2016-10-31
Root cause
Under Investigation by firm
Terminated
2018-01-09
Distribution
US Nationwide Distribution.
Recall number
Z-0888-2017
Recall status
Terminated
CDRH record ID
150987
FDA record created
2016-12-22
Recall event number
75649

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗