Skip to content

Public FDA record

Z-0890-2013: Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample, Not for

Recall record Z-0890-2013 for Oculus Innovative Sciences Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0890-2013
Company in FDA record
Oculus Innovative Sciences Inc
FDA classification date
Feb 26, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.

Source facts

Action
Oculus sent an Urgent Medical Device Recall letter dated February 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue distribution and segregate the affected product. Immediately remove the affected lots of Recall product from the inventory of sales representative, sub contracted distribution and or fulfillment centers. Segregate the affected product in a secure location for return or destruction. If the product was further distributed customers were instructed to notify healthcare professionals of this recall. Customers were also instructed to completed and return the Business Reply Card and fax to 1-415-462-5163. Once the firm receives the affected product the customer would be contacted by an Oculus Representative and provided a Returns Goods Authorization (RGA) number. For questions customers were instructed to call 1-707-559-2445. For questions regarding this recall 707-559-7191, ext 14.
Initiated
2013-02-12
Root cause
Incorrect or no expiration date
Terminated
2013-11-25
Distribution
Nationwide Distribution - One distributor in PA
Recall number
Z-0890-2013
Recall status
Terminated
CDRH record ID
115812
FDA classified
2013-02-26
Recall event number
64316

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗