General Electric Medical Systems Information Technology
FDA classification date
Jun 6, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot .
Source facts
Action
GE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07 that explained the short and long term solutions and that the company would be providing a software update that will rectify the issue specific to CIC Pro v4.1 software.
Initiated
2007-01-24
Root cause
Other
Terminated
2011-12-10
Distribution
Nationwide
Recall number
Z-0891-2007
Recall status
Terminated
CDRH record ID
50585
FDA classified
2007-06-06
Recall event number
37309