There is a potential for failure of a lead screw drive system or image intensifier lead screw drive assembly, if normal maintenance inspections are not followed correctly.
Source facts
Action
Varian sent an URGENT MEDICAL DEVICE CORRECTION letter via certified mail receipt on January 19, 2010, to all direct consignees. The letter identified the product, the problem, and the action the consignee should take.
Consignees were instructed to have maintenance personnel perform the recommended measurements as described in the Varian Customer Technical Bulletin.
Ensure that all measurements made as part of the procedure are recorded in the 'wear table' that is part of the procedure.
Previous wear measurements should be compared to new measurements and any large increase in wear should be investigated immediately.
Pay special attention to the "Recommendations based on wear measurements" table.
Ensure that new wear measurements comply with the operating parameters described in the table.
For any questions consignees are to contact Varian Medical Systems Customer Support Services at one of the locations described in the CTB.
For questions regarding this recall call (650) 424-5731.
Initiated
2010-01-19
Root cause
Other
Terminated
2012-02-24
Distribution
Worldwide Distribution including USA, Australia, New Zealand, Europe, Southeast Asia, China, Japan, Latin America, and Canada.
Recall number
Z-0892-2011
Recall status
Terminated
CDRH record ID
93282
FDA classified
2011-01-11
Recall event number
56350