- Recall classification
- Class II
Reason stated in the source
Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
Source facts
- Action
- Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
- Initiated
- 2014-11-20
- Root cause
- Device Design
- Terminated
- 2018-01-10
- Distribution
- US Nationwide Distribution
- Recall number
- Z-0892-2015
- Recall status
- Terminated
- CDRH record ID
- 131596
- FDA classified
- 2014-12-30
- Recall event number
- 69821