Open original FDA record ↗Recall record
- FDA identifier
- Z-0892-2018
- Company in FDA record
- Bausch & Lomb Inc Irb
- FDA record creation date
- Mar 5, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The complaints database was reviewed and there have been an atypical volume of dull blades complaints for various lots. out of an abundance of precaution, Bausch + Lomb is voluntarily recalling all product codes SKU manufacturing without the electro-polish control and penetration force testing.
Source facts
- Action
- On March 3, 2017 Bausch + Lomb , distributed Urgent Medical Device Recall notices and Acknowledgement Forms to customers via FedEx. Customers were advised to: 1. Please review your inventory and quarantine all unused (full and partial) boxes of LaserEdge" knives (6/Box or individual packaged knives) affected lots. 2. Complete the enclosed Medical Device Voluntary Recall Acknowledgement Form and contact Bausch + Lomb to obtain a Return Material Authorization Number (RMA) and arrange for a pickup of the identified product. You can contact the Bausch + Lomb Surgical Customer Service team by calling 1-800-338-2020, Customer Service (option 2), Product Returns and Adverse Event Reporting (option 2). Please contact the Bausch + Lomb Surgical Customer Service team with any questions or concerns regarding this process: 1-800-338-2020.
- Initiated
- 2017-03-03
- Root cause
- Process design
- Terminated
- 2018-06-25
- Distribution
- US Nationwide Distribution
- Recall number
- Z-0892-2018
- Recall status
- Terminated
- CDRH record ID
- 160023
- FDA record created
- 2018-03-05
- Recall event number
- 76874