Potential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures. The increase in temperature could cause a reduction of power, causing the solenoid in the valve to function sporadically.
Source facts
Action
Notified consignees by letter on 04/04/2005.
Initiated
2005-04-04
Root cause
Other
Terminated
2007-11-27
Distribution
Nationwide, Military medical center, Canada, and Czech Republic
Recall number
Z-0893-06
Recall status
Terminated
CDRH record ID
42764
FDA classified
2006-05-23
Recall event number
33938