- FDA identifier
- Z-0893-2022
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Apr 7, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a quality issue with the three bolts that hold the CT balancing weights, which could create a safety issue.
Source facts
- Action
- Customer Advisory Notice Letter will be provided to each impacted consignee. Depending on the customer s capabilities, Siemens provides the letters via mail, e-mail, or hand delivery by a service engineer. The receipt of the letter will be documented, and additional attempts will be made until such receipt is acknowledged. The safety advisory letter informs customers: You may continue to use your system ensuring to follow all instructions in the Operator Manual. If any changes in system vibrations or noises are detected, then activate the Emergency Stop (E-Stop), discontinue use of the system immediately and contact your service engineer.
- Initiated
- 2022-03-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2023-04-14
- Distribution
- Distribution OUS only to El Salvador, France, Germany, Japan, Taiwan, and UK.
- Recall number
- Z-0893-2022
- Recall status
- Terminated
- CDRH record ID
- 192435
- FDA record created
- 2022-04-07
- Recall event number
- 89790