Open original FDA record ↗Recall record
- FDA identifier
- Z-0895-2013
- Company in FDA record
- Orthofix, Inc
- FDA classification date
- Feb 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.
Source facts
- Action
- Orthofix sent an Urgent Product Recall Notification letter dated August 6, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue distribution of devices, identify the affected and remove from their inventory, and returned to the firm. Credit will be issued upon receipt of the returned product. Customers were asked to take the actions listed in the enclosed Product Return Instructions immediately. Customers with questions should contact Orthofix Customers Service by telephone at 800-266-3349 or 214-937-2047. For questions regarding this recall call 214-937-2509.
- Initiated
- 2012-08-06
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-05-17
- Distribution
- Worldwide Distribution - USA (nationwide) Internationally to Australia, Hong Kong, Saudi Arabia, Korea, Turkey, Greece, Finland, South Africa, Spain, Singapore, Switzerland, Brazil, Germany, France, Japan, Netherlands, New Zealand, Israel, Canada and Mexico
- Recall number
- Z-0895-2013
- Recall status
- Terminated
- CDRH record ID
- 111748
- FDA classified
- 2013-02-28
- Recall event number
- 62804