Philips Medical Systems North America Co. Phillips
FDA classification date
Jun 16, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Automated analysis of vascular obstruction tool in software gives unreliable calibration and analysis results when using unprocessed images acquired on the Philips Allura FD10.
Source facts
Action
On 4/29/05 the firm mailed a letter, via certified return receipt. to their customers. The letter notified users of the problem and offered a work around. Consignees were advised they would be contacted by a representative to upgrade the unit with new software.
Initiated
2005-04-29
Root cause
Other
Terminated
2008-05-09
Distribution
The units were distributed to 18 medical centers thorughout the US.
Recall number
Z-0896-05
Recall status
Terminated
CDRH record ID
38888
FDA classified
2005-06-16
Recall event number
31997