Open original FDA record ↗Recall record
- FDA identifier
- Z-0896-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Jan 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software safety guard measures on the device malfunctioned to prevent the temperature from going over its limits when the device is used.
Source facts
- Action
- Field notification in the form of a Customer Safety Advisory Notice was initiated on 09/30/2009 by the firm, and the notice was mailed to all affected customers Siemens sent a CUSTOMER SAFETY ADVISORY NOTIFICATION letter on September 30, 2009. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconnect and reconnect the V5Ms transducer between uses until their local Customer Service Engineer installs the software update to their system. For any questions customers were to contact their local service support person. For questions regarding this recall call 650-694-5398.
- Initiated
- 2009-09-30
- Root cause
- Software design
- Terminated
- 2012-06-18
- Distribution
- Worldwide Distribution - United States (nationwide) and the countries of Australia, Austria, Azerbaijan, Brazil, Belarus, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Guatemala, Croatia, India, Indonesia, Italy, Jordan, Japan, Latvia, Malaysia, Mexico, Moldova, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey, Ukraine, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
- Recall number
- Z-0896-2011
- Recall status
- Terminated
- CDRH record ID
- 89055
- FDA classified
- 2011-01-12
- Recall event number
- 54718