Open original FDA record ↗Recall record
- FDA identifier
- Z-0896-2013
- Company in FDA record
- Orthofix, Inc
- FDA classification date
- Feb 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.
Source facts
- Action
- Orthofix sent an Urgent Medical Device Recall Notification dated February 1, 2013 to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to identify and remove the affected products from inventory immediately and follow the actions listed in the enclosed Product Return Instructions. For questions or concerns contact Regulatory Affairs at 214-937-2100.
- Initiated
- 2013-02-01
- Root cause
- Device Design
- Terminated
- 2013-07-01
- Distribution
- Worldwidewide Distribution - USA Nationwide including Puerto Rico and the countries of: Australia, Lebanon, Germany, Spain, Sweden and Israel.
- Recall number
- Z-0896-2013
- Recall status
- Terminated
- CDRH record ID
- 115728
- FDA classified
- 2013-02-28
- Recall event number
- 64265