Open original FDA record ↗Recall record
- FDA identifier
- Z-0896-2015
- Company in FDA record
- Hansen Medical Inc
- FDA classification date
- Dec 30, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.
Source facts
- Action
- Hansen Medical sent an Urgent Notice - Field Safety Notice letters to US customers on December 4, 2014 via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and fax back the attached form to Hansen medical at 650-404-2773. For further information call 650-404-5841.
- Initiated
- 2014-12-04
- Root cause
- Device Design
- Terminated
- 2015-09-03
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Italy, Czech Republic, Russia, Spain, Singapore, France, Germany, Denmark, Netherlands and Australia.
- Recall number
- Z-0896-2015
- Recall status
- Terminated
- CDRH record ID
- 131725
- FDA classified
- 2014-12-30
- Recall event number
- 69855