If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
Source facts
Action
The firm notified their consignees of the issue by Fed Ex on 01/20/2016. The firm plans to coordinate with the medical facilities for inspection of the endoscopes.
Initiated
2016-01-20
Root cause
Nonconforming Material/Component
Terminated
2016-10-13
Distribution
Distributed US (nationwide) and in Australia.
Recall number
Z-0896-2016
Recall status
Terminated
CDRH record ID
143130
FDA record created
2016-02-25
Recall event number
73088