- FDA identifier
- Z-0896-2022
- Company in FDA record
- DRG International, Inc.
- FDA record creation date
- Apr 7, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ODs of the standards too low(< 0.6 ODmax) may cause invalid runs, delayed patient results are possible
Source facts
- Action
- DRG issued letter via email on 12/28/21. Letter states reason for recall, health risk and Action to be taken by distributors and end-users: Please check your inventory for the respective lot, stop using these kits immediately and promptly quarantine these products. Only if you faced above mentioned issue of SO OD <1.2, please ask for replacement of the kits that were affected and provide your data. All unused kits of the affected lot will be replaced.Please reply immediately upon checking your stock/use to this notice to confirm. All replies should be received at once OR by January 7th, 2022. We sincerely apologize for the inconvenience this may have caused you and/or your Customer. Reply to email: qa@drg-international.com and copy drgsupport@drginternational.com.
- Initiated
- 2021-12-21
- Root cause
- Under Investigation by firm
- Terminated
- 2024-02-12
- Distribution
- NY
- Recall number
- Z-0896-2022
- Recall status
- Terminated
- CDRH record ID
- 192206
- FDA record created
- 2022-04-07
- Recall event number
- 89680