Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.
Source facts
Action
Field modification instructions were issued on 3/24/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to the hospitals.
Initiated
2004-03-24
Root cause
Other
Terminated
2006-07-24
Distribution
The units were distributed nationwide in the United States and worldwide.