Open original FDA record ↗Recall record
- FDA identifier
- Z-0897-2011
- Company in FDA record
- Phillips Nuclear Medicine
- FDA classification date
- Jan 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A problem has been identified with some upper patient pallets on Precedence SPECT/CT System, which may reduce its load bearing capacity and may result in its cracking. If this were to happens, could pose a risk to patient.
Source facts
- Action
- The firm, Philips Healthcare, sent an "Urgent - Field Safety Notice" dated January 26, 2009 to all consignees/customers. The notice described the product, problem and action to be taken by customers and by Philips. The customers were instructed to reference their system's System Operation and System information Manuals regarding loading and unloading of patient; monitoring the patient during a study ,and careful monitoring during unsupported (cantilevered) table translation. Philips has ordered replacement parts and they will have a Philip Healthcare Field Service Engineer contact and visit the customers site within two months to replace the affected upper patient pallet. If you need any further information or support concerning this issue, please contact your local Philips representative at: Customer Care Center (1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: and Oncology or your local Philips Healthcare office.
- Initiated
- 2009-01-26
- Root cause
- Process change control
- Terminated
- 2011-01-13
- Distribution
- International distribution only: Mexico and Saudi Arabia.
- Recall number
- Z-0897-2011
- Recall status
- Terminated
- CDRH record ID
- 79567
- FDA classified
- 2011-01-12
- Recall event number
- 51274