Z-0897-2022: Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon dioxid

Recall record Z-0897-2022 for Maquet Medical Systems USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0897-2022
Company in FDA record
Maquet Medical Systems USA
FDA record creation date
Apr 8, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.

Source facts

Action
Getinge issued Urgent Medical Device Correction Letter dated 2/18/22 via FedEx 2 day service with signature required. . Letter states reason for recall, health risk and action to take: A general decommissioning of the affected Cardiohelp-i is not required, with the consideration of the following: 1. Please examine your inventory immediately to determine if you have received any Cardiohelp-i products with one of the affected serial numbers. 2. If you have a Cardiohelp-i affected by this field correction, please examine your inventory immediately to locate all Transport Guards used with Cardiohelp-i products. Please ensure that the Transport Guard can be opened and that the cylindrical pin is not loose prior to using the device. If so, the device can continue to be used. If your Transport guard cannot be opened or if the cylindrical pin is loose, please remove from use and place in a quarantine location to avoid future use. This activity should be performed prior to every use until the field correction activities have been completed at your facility by a Getinge Representative. 3. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 5). A Getinge Service Technician will contact the person you list on the Customer Response Form to schedule service for an inspection of your device. Return the completed form to Maquet Cardiopulmonary GmbH /Getinge by emailing a scanned copy to Transportguardpin2022.act@getinge.com or by faxing the form to 1(888) 722-5375. If your Transport Guard displays the issue identified in this notification, it will be replaced at no cost to your facility. 4. Whether or not you have affected product(s) with the serial numbers listed in this notice, please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 5) to acknowledge that you have received this notification, per the instruction in number 3 above. 5. You can also arrange a visit by a Getinge trained or au
Initiated
2022-02-18
Root cause
Process control
Distribution
US Nationwide Distribution: FL, KY, NH, NJ, NY, PA, TX, VA Foreign: Belgium, Canada, Germany, Hungary, Luxemburg, Malaysia, Mexico, Poland, Slovakia, South Africa, South Korea, and Taiwan.
Recall number
Z-0897-2022
Recall status
Open, Classified
CDRH record ID
192339
FDA record created
2022-04-08
Recall event number
89745

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗