Reason stated in the source
Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient.
Source facts
- Action
- GEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.
- Initiated
- 2006-03-06
- Root cause
- Other
- Terminated
- 2008-02-24
- Distribution
- Nationwide including Puerto Rico and OUS to include: Australia, Belgium, Brazil, Bulgaria, Byelarus, Canada, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Honduras, Hungary, Indonesia, Iraq, Ireland, Israel, India, Iran, Italy, Jordan, Japan, Republic of Korea, Lebanon, Lithuania, Martinique, Mexico, Morocco, Netherlands, Norway, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Republic of China, Thailand, Tunisia, Turkey, United Kingdom, Venezuela
- Recall number
- Z-0898-06
- Recall status
- Terminated
- CDRH record ID
- 45084
- FDA classified
- 2006-05-24
- Recall event number
- 34758