Open original FDA record ↗Recall record
- FDA identifier
- Z-0898-2018
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Mar 5, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain Auriga consoles have incorrectly assembled trigger wire components. The trigger wire functions in the start-up of the console. An incorrectly assembled trigger wire may result in error codes or failure to start the console.
Source facts
- Action
- Urgent Medical Device Field Correction letters, dated September 2017, were sent to distributors and medical facilities with instructions for acknowledging the notice and making corrections. The letter identified the affected device and the reason for the correction. Customers were to check their inventory, segregate, and cease distribution for the affected devices. Customers were to complete and return the acknowledgment form. **NO PRODUCT IS BEING RECALLED AND YOU ARE NOT REQUIRED TO RETURN PRODUCT TO BOSTON SCIENTIFIC**
- Initiated
- 2017-09-22
- Root cause
- Process design
- Terminated
- 2018-08-14
- Distribution
- International Distribution Only -- Australia, Vietnam, France, Germany, Italy, Spain, and Romania.
- Recall number
- Z-0898-2018
- Recall status
- Terminated
- CDRH record ID
- 161329
- FDA classified
- 2018-03-05
- Recall event number
- 79096