Open original FDA record ↗Recall record
- FDA identifier
- Z-0898-2022
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA record creation date
- Apr 8, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sterility failure of the exterior of the syringe which may impact the device's ability to be laid on a sterile field.
Source facts
- Action
- The recalling firm issued letters dated 3/4/2022 via regular mail informing the customer there was a sterility failure of the exterior of the syringe and in the event a syringe from the affected lot was placed on a sterile field, the sterile field may become compromised. The required actions were to immediately check their stock for the affected lot number, quarantine it, and then destroy it. The enclosed destruction form was to be completed listing the quantity of affected product destroyed, and once the form was returned, they would receive credit. If the product has been transferred to another individual or location, they were to notify them of the recall.
- Initiated
- 2022-03-04
- Root cause
- Material/Component Contamination
- Terminated
- 2023-03-20
- Distribution
- US Nationwide Distribution was made to KY, OK, and TX. There was no foreign/government/military distribution.
- Recall number
- Z-0898-2022
- Recall status
- Terminated
- CDRH record ID
- 192443
- FDA record created
- 2022-04-08
- Recall event number
- 89795