Open original FDA record ↗Recall record
- FDA identifier
- Z-0899-2012
- Company in FDA record
- Angiodynamics Worldwide Headquarters
- FDA classification date
- Jan 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The red luer on the Morpheus CT PICC (Peripherally Inserted Central Catheter) has the potential to crack during use of the product.
Source facts
- Action
- AngioDynamics, Inc. sent an Urgent - Medical Device Recall letter dated January 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify and segragate the recalled affected product. Complete the enclosed Morpheus Ct PICC Recall Reply Form and fax it to the attention of the Morpheus CT PICC Recall Coordinator at 518-798-1360. If customers do not have any affected product remaining in their possession they should fill out the attached form noting zero quantity to be returned and fax the form to 518-798-1360. If customers distributed the affected product, they should inform their customers of this recall action immediately and have them return the recalled units to them. For any questions regarding this recall call 1-800-772-6446 or e-mail customerservice@angiodynamics.com.
- Initiated
- 2012-01-09
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-02-05
- Distribution
- Nationwide
- Recall number
- Z-0899-2012
- Recall status
- Terminated
- CDRH record ID
- 106934
- FDA classified
- 2012-01-26
- Recall event number
- 60909