Open original FDA record ↗Recall record
- FDA identifier
- Z-0899-2017
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Dec 22, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.
Source facts
- Action
- Radiomet4er sent a letter dated November 30, 2016, to all affected customers via email with a request for customers to complete a return form. Customers/Users were advised to not use the "more" button to expand patient results when viewing them in the AQURE patient view. Customers/Users Radiometer Representative will contact them to schedule a visit of a remote session. During the visit or remote session the representative will run a script on the AQURE system, which serve to remove the misalighnment possibility. It is noted in the letter to customers that expanding the results by using the More button will not affect the stored results. If customers are not the end-user of the affected product ensure that the notification letter is distributed to the final end-user. If any questions regarding the notification letter contact Radiometer America Technical Support at 1-800-736-0600 opt.1. For questions regarding this recall call 440-871-8900.
- Initiated
- 2016-11-23
- Root cause
- Software design
- Terminated
- 2017-02-07
- Distribution
- Worldwide Distribution - US including MI and NY and Internationally to England, Switzerland, and United Arab Emirates.
- Recall number
- Z-0899-2017
- Recall status
- Terminated
- CDRH record ID
- 151434
- FDA classified
- 2016-12-22
- Recall event number
- 75836