McKesson Provider Technologies - Medical Imaging Group
FDA classification date
Jan 13, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
When importing ECG files from Mortara ECG carts using DICOM interface, the Horizon Cardiology ECG Management System does not display the amplitudes of the waveform properly.
Source facts
Action
McKesson Medical sent an advisory notice letter dated June 16, 2010, to all affected customers. Advisory notice ADV01646 was distributed to 13 affected clinical sites from August 30, 2010 to September 28, 2010. The letters identified the product that is affected, the circumstances under which patient safety issue may occur, the risk, and the immediate actions that need to be taken along with a temporary fix.
Customers were instructed to use the ECG cart printout to verify results of procedures acquired by Mortara ECG carts.
Users should be aware of the problem when referrring to past ECG procedures and when performing a visual comparison with ECG procedures acquired by Mortara ECG carts.
Customers should disable the Glasgow Serial Comparison for ECG procedures.
McKesson Support would contact the customer to coordinate the deployment of the software update on all the affected applications and Horizon Cardiology client applications.
For any questions regarding this recall, customers should call the Customer Support department at 1-866-777-0202 and speak to their Support Manager.
Initiated
2010-06-16
Root cause
Device Design
Terminated
2012-08-17
Distribution
Nationwide Distribution including AZ, CA, FL, GA, MA, PA, UT, and WI.
Recall number
Z-0900-2011
Recall status
Terminated
CDRH record ID
96241
FDA classified
2011-01-13
Recall event number
57436