Z-0900-2022: LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and

Recall record Z-0900-2022 for Kaneka America Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0900-2022
Company in FDA record
Kaneka America Corp
FDA record creation date
Apr 9, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.

Source facts

Action
On 02/17/2022 a recall notification letter with revised labeling was emailed regarding contraindication in patients who are being treated with angiotensin-converting enzyme (ACE) inhibitors. Customers are to locate, destroy and replace all operator's manuals, patient guides, and IFUs with those attached to the recall notice, ensuring they are distributed to all end users. All device model, lot, and UDI numbers for systems and disposables are affected. Submit the Acknowledgement and Receipt Form to Kaneka Medical America, LLC via email. If you have any question or concerns about this recall please contact Kaneka Medical America LLC Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time, at: Phone: (212) 705-4355, Email: Ahmad.Al-Sattari@kaneka.com
Initiated
2022-02-17
Root cause
Error in labeling
Terminated
2024-05-06
Distribution
US only: CA, NH, CT, MA, FL, DE, NY, NJ, PA, OH, MD, VA, NC, SC, GA, MS, TN, TX, IN, AZ, MI, WI, MN, IL, MO, NE, CO, WA, OR, UT, KS, OK, AL, ME
Recall number
Z-0900-2022
Recall status
Terminated
CDRH record ID
192323
FDA record created
2022-04-09
Recall event number
89735

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗