Open original FDA record ↗Recall record
- FDA identifier
- Z-0901-2014
- Company in FDA record
- Merit Medical Systems, Inc.
- FDA classification date
- Jan 30, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Merit Medical Systems, Inc. is voluntarily recalling one lot (B507171) of RAD BOARD RB 100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There
Source facts
- Action
- Merit Medical Systems, Inc.sent a letter voluntarily recalling one lot (B507171) of RAD BOARD" RB100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There have been no reports of patient harm or injury from Merit customers as a result of this issue. Merit has chosen to remove this lot from the field by replacing the unlabeled RAD BOARDs. Merit requests that customers stop using these devices immediately until the product has been replaced by their Merit sales representative. Customers with questions were instructed to call 1-801-208-4344 or 1-801-208-4365. For questions regarding this recall call 801-208-4301.
- Initiated
- 2013-12-18
- Root cause
- Labeling Change Control
- Terminated
- 2014-02-13
- Distribution
- Worldwide Distribution - USA including NJ, TX, WV, MN, GA, and Internationally to Mexico and Japan.
- Recall number
- Z-0901-2014
- Recall status
- Terminated
- CDRH record ID
- 124977
- FDA classified
- 2014-01-30
- Recall event number
- 67206