Skip to content

Public FDA record

Z-0901-2022: IntraSight Mobile IVUS Systems, Model: 797415, Part numbers: 300002018101, 300002018102

Recall record Z-0901-2022 for Volcano Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0901-2022
Company in FDA record
Volcano Corp
FDA record creation date
Apr 9, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Multi-Modality Touch Screen Modules (MM-TSM), which are bedside/secondary control units, used as part of Mobile IVUS systems, are missing the following labeling information: Unique Device Identifier, Legal Manufacturer Information, Part Number, Rx Symbol, and Model Reference Information.

Source facts

Action
On 01/24/22, correction notices were sent to customers who were informed that the recalling firm will notify all affected customers and arrange for a Field Service Engineer (FSE) to replace the MM-TSM free of charge. In addition, customers were asked to do the following: 1) Distributors or customers who have forwarded affected devices to other end users were asked to, send a copy of the correction notification to any customer to whom they have distributed affected devices to. 2) Customers were asked to complete and return customer response forms. 3) Customers with additional questions were encouraged to contact their local firm representative or contact IGTD Customer Service: Phone: 1-800-228-4728, Option 2, Email: igtd.remotesupport@philips.com, Hours of Operation: Monday - Friday 8:00AM - 5:00PM PST
Initiated
2022-01-24
Root cause
Error in labeling
Distribution
US: CO, GA, MA, TX, AZ, NY, PA, CA, FL, MO, OR, MD, NC, TN OUS: JP, DE, GB, NL, IT
Recall number
Z-0901-2022
Recall status
Open, Classified
CDRH record ID
192183
FDA record created
2022-04-09
Recall event number
89677

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗