Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.
Source facts
Action
Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]
Initiated
2003-04-24
Root cause
Other
Terminated
2003-09-25
Distribution
United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela
Recall number
Z-0902-03
Recall status
Terminated
CDRH record ID
26689
FDA classified
2003-05-29
Recall event number
25899