Open original FDA record ↗Recall record
- FDA identifier
- Z-0902-2013
- Company in FDA record
- Salyer Prn Biomedical
- FDA classification date
- Mar 1, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Concerns that the addition of the UPS Battery Backup to the Century Console Base Shelf did not comply with the FDA premarket notification regulations.
Source facts
- Action
- An Urgent: Medical Device Recall letter, dated 1/31/13, was sent to all the customers that received the Century Console Base Shelf option with the Linear/1.5k 115v UPS Battery Backup for the Century Heart Lung Machine. The letter informed the customers that Salyer PRN Biomedical will have a Salyer Biomedical technician return the customers' Century Console Bases to their original configuration without the Shelf option and restore the Linear/1.5k UPS Battery Backup to its original condition as a stand-alone unit with its own wheel base for ease of use. A refund or credit will be issued for the return of the product. Customers are instructed to contact Jeff Poland at (314) 497-5232.
- Initiated
- 2013-01-31
- Root cause
- No Marketing Application
- Terminated
- 2013-04-15
- Distribution
- Distributed in the states of MA,CA, and WI.
- Recall number
- Z-0902-2013
- Recall status
- Terminated
- CDRH record ID
- 115964
- FDA classified
- 2013-03-01
- Recall event number
- 64383