HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a
false high or false low results for HbA1c and false high results for magnes
Source facts
Action
A recall letter was sent on 11/13/14 to all customers who purchased the ABX PENTRA 400 Clinical Chemistry Analyzer. The letter informs the customer of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact their local HORIBA Medical representative.
Initiated
2014-11-13
Root cause
Package design/selection
Terminated
2015-03-19
Distribution
Nationwide in US
Recall number
Z-0902-2015
Recall status
Terminated
CDRH record ID
131529
FDA classified
2015-01-02
Recall event number
69773