Skip to content

Public FDA record

Z-0902-2017: Specimen Gate Laboratory 5002-0180, Specimen Gate eReports 5002-0370, Software Version 1.0 Specimen Gate

Recall record Z-0902-2017 for PerkinElmer Health Sciences, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0902-2017
Company in FDA record
PerkinElmer Health Sciences, Inc.
FDA classification date
Dec 23, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Issue may cause eReports to consume an excessive amount of disk space causing the software to stop operating.

Source facts

Action
PerkinElmer Health Sciences issued a Medical Device Correction letter dated December 1, 2016, to all affected customers email facilitated through our PerkinElmer Software Services group located in Akron Ohio. Software Services will be asked to make sure the customers receive the Recall Letter and implement the recommended corrective measures. Actions to be taken A field correction of the issue is recommended. PerkinElmer will correct the issue with an update installed through remote connection. You will be contacted by your PerkinElmer representative to make the necessary arrangements for implementing the update. In the interim period, please do not attempt to login multiple times in quick succession. Allow the software time to respond to first login command. Please contact your local PerkinElmer representative or Specimen Gate Support for further information. Other Information Please inform those affected in your organization accordingly. To comply with regulatory requirements we request that you complete the enclosed response form and return it by fax to number 1-330-825-8520 (United States) /+358 2 2678 357 (outside United States) or as scanned by e-mail to TurkuQMresponse@perkinelmer.com as soon as possible, but not later than December 22, 2016. For questions regarding this recall call 781-663-6900.
Initiated
2016-12-01
Root cause
Software design
Terminated
2017-05-01
Distribution
Worldwide Distribution - US including CO, FL, GA, IL, MA, MI, NV, NJ, OH, SC, AZ, KY and Internationally to Italy, Ireland, UK, Spain, and Peru
Recall number
Z-0902-2017
Recall status
Terminated
CDRH record ID
151383
FDA classified
2016-12-23
Recall event number
75817

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗