Open original FDA record ↗Recall record
- FDA identifier
- Z-0903-2015
- Company in FDA record
- DePuy Orthopaedics, Inc.
- FDA classification date
- Jan 5, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
Source facts
- Action
- DePuy Synthes sent an Urgent Information letter dated November 10, 2014, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to complete the Hospital/User Facility Acknowledgement Car and return to their Deputy Orthopaedics Sales Consultant or fax to 574-371-4939. Customer swere also asked to retain a copy of the complete Acknnowlegement Card in their files along with the notice. Customers with questions were instructed to call 574-371-4917 (M-F; 8 a.m. 5 p.m. EDT). For questions regarding this recall cal 574-371-4917.
- Initiated
- 2014-11-10
- Root cause
- Device Design
- Terminated
- 2015-12-08
- Distribution
- Worldwide Distribution: US (nationwide) including P.R. except AK and RI; and Internationally to: Chile, Canada, Australia, Austria, Belgium, China, France, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK).
- Recall number
- Z-0903-2015
- Recall status
- Terminated
- CDRH record ID
- 132180
- FDA classified
- 2015-01-05
- Recall event number
- 70043