Z-0903-2022: Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-001

Recall record Z-0903-2022 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0903-2022
Company in FDA record
Merit Medical Systems, Inc.
FDA record creation date
Apr 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.

Source facts

Action
On 03/01/22, recall notices were mailed to customers wo were informed to take the following actions: 1. Quarantine and discontinue use, and distribution of affected devices. 2. Ensure that applicable personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure the recall notice is immediately shared with them and note the quantity distributed on the Customer Response Form. 4. Fill out, scan, and email the completed Customer Response Form to Customer Service at response@merit.com. All affected product shipped to you must be accounted for on the Customer Response Form. 5. Immediately return all affected lots to the recalling firm. Customers with additional questions are encouraged to contact a sales representative or the firm s customer service at RESPONSE-EMEA@merit.com, 1-800-356-3748, Hours: 6 am to 6 pm MST, Mon-Fri. Recall notices were also sent to Sales representatives who were asked to contact their affected accounts and assist with product return and completion of the Customer Response Form. On 03/21/22, recall notices and response forms including an expanded scope were sent to customers informing them of additional distributed amounts of affected devices.
Initiated
2022-03-01
Root cause
Component change control
Distribution
US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO, LA, OR, MT, MA, MI, NE, WV, KS, MS, HI, IA, NH, ME, AL, DC, DE, NM, OH, GA, MD, TN, UT, RI, SC, OK, WA, ID, ND, SD, NV, VT. OUS: Poland, Germany, France, United Kingdom, Belgium, Korea, Republic of, Austria, Canada, Spain, Luxembourg, Italy, Switzerland, Czech Republic, Netherlands, Japan, Honduras, Colombia, Greece, Ireland
Recall number
Z-0903-2022
Recall status
Open, Classified
CDRH record ID
192626
FDA record created
2022-04-11
Recall event number
89815

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗