Open original FDA record ↗Recall record
- FDA identifier
- Z-0904-06
- Company in FDA record
- Joerns Healthcare Inc.
- FDA classification date
- May 25, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A field correction is being conducted on select models of Hoyer Patient Lifts because of hydraulic jack problems that have resulted in patient injury. The possibility exists that the mounting stud located in the bottom of the hydraulic jack could become loose allowing the jack to disengage from the mast of the lift.
Source facts
- Action
- A Technical Service Bulletin was distributed along with an Urgent Medical Device Correction letter providing details for checking the mounting stud. UPDATE: Sunrise Medical sent a Second Notice Urgent Medical Device Correction letter, in follow-up to the January 2006 notice, to Dealer/Distributors starting 04/26/06. This Notice updates customers as to what Sunrise''s field correction will be for the Hoyer Model Hydraulic Patient Lifts. The Notice states that use of the affected devices should be discontinued until the lift has been inspected and repaired. Dealers/Distributors that have distributed the product are to notify the user to discontinue use, if possible. If the lift needs to be used then the Dealer/Distributor is to do an interim maintenance.
- Initiated
- 2006-01-06
- Root cause
- Other
- Terminated
- 2008-03-02
- Distribution
- Nationwide, Canada, Asia, South America
- Recall number
- Z-0904-06
- Recall status
- Terminated
- CDRH record ID
- 43199
- FDA classified
- 2006-05-25
- Recall event number
- 34153