Z-0904-06: Hoyer Hydraulic Patient Lifters model C-CB-L2, C-HLA, C-HLA-2, HML-400, C-HLA-2T and C-HLA-1. Distribute

Recall record Z-0904-06 for Joerns Healthcare Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0904-06
Company in FDA record
Joerns Healthcare Inc.
FDA classification date
May 25, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A field correction is being conducted on select models of Hoyer Patient Lifts because of hydraulic jack problems that have resulted in patient injury. The possibility exists that the mounting stud located in the bottom of the hydraulic jack could become loose allowing the jack to disengage from the mast of the lift.

Source facts

Action
A Technical Service Bulletin was distributed along with an Urgent Medical Device Correction letter providing details for checking the mounting stud. UPDATE: Sunrise Medical sent a Second Notice Urgent Medical Device Correction letter, in follow-up to the January 2006 notice, to Dealer/Distributors starting 04/26/06. This Notice updates customers as to what Sunrise''s field correction will be for the Hoyer Model Hydraulic Patient Lifts. The Notice states that use of the affected devices should be discontinued until the lift has been inspected and repaired. Dealers/Distributors that have distributed the product are to notify the user to discontinue use, if possible. If the lift needs to be used then the Dealer/Distributor is to do an interim maintenance.
Initiated
2006-01-06
Root cause
Other
Terminated
2008-03-02
Distribution
Nationwide, Canada, Asia, South America
Recall number
Z-0904-06
Recall status
Terminated
CDRH record ID
43199
FDA classified
2006-05-25
Recall event number
34153

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗