Z-0904-2021: Philips Respironics V60 Ventilator Part Number 1053615 - Product Usage: intended to be used only with var

Recall record Z-0904-2021 for Respironics California, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0904-2021
Company in FDA record
Respironics California, LLC
FDA classification date
Jan 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.

Source facts

Action
On 11/02/2020, the firm sent an "URGENT-MEDICAL DEVICE CORRECTION" Notification Letter via priority mail informing customer that the V60 Ventilators may experience two potential issues with the navigation ring interface (see table below for additional information). 1. The navigation ring may become inoperative or begin operating intermittently. 2. The navigation ring may not behave as expected, making it difficult for the user to change ventilator settings via the touchscreen. and if clinicians accept an incorrect setting and leaves the patient on the ventilator without responding to the resulting alarms, the patient could sustain serious injuries. Customers are instructed: 1. When cleaning V60 ventilators, do not allow cleaning agents to pool on the ventilator surfaces. This is especially important on the device interface. 2. Check setting changes before and after accepting the changes. 3. If the navigation ring becomes inoperative or operates intermittently during clinical use, use the touchscreen for all settings changes and navigation, and contact Philips to schedule service following clinical use. 4. If the navigation ring interferes with your ability to adjust therapy settings via touchscreen, remove the ventilator from service and contact Philips for repair. and to please review the following sections of the user manual: 1. Chapter 10: Care and maintenance. To prevent possible damage to the ventilator when cleaning, use only those cleaning and disinfecting agents listed, and do not drip or spray any liquids directly onto any ventilator surface. 2. Chapter 1: Warnings, cautions, and notes. During clinical use of the ventilator, monitor patients for changes in respiratory status with each ventilator setting adjustment. Use an external oxygen monitor to verify oxygen concentration, and have an alternative means of ventilation available whenever a ventilator is in use. Customers are also instructed to complete, sign and return the Acknowledgement and Receipt
Initiated
2020-11-02
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide including Guam and Puerto Rico; Afghanistan, Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Ethiopia, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, ST. Pier Miquel, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, Wallis Futuna, and Zambia.
Recall number
Z-0904-2021
Recall status
Open, Classified
CDRH record ID
185071
FDA classified
2021-01-22
Recall event number
86697

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗