If a customer enters a dilution factor for Sample B prior to the results from Sample A being transfered to the workstation, the predilute multiplication factor will be applied erroneously to the Sample A's results once they are received at the workstation.
Source facts
Action
A Product Corrective Action Letter was mailed to all customers who currently have the instrument in thier lab. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occuring. The letter was sent by US Mail on March 31, 2004.
Initiated
2004-03-31
Root cause
Other
Terminated
2012-04-30
Distribution
Nationwide and Canada
Recall number
Z-0905-04
Recall status
Terminated
CDRH record ID
32664
FDA classified
2004-07-20