Skip to content

Public FDA record

Z-0905-06: Medtronic Model 37711 Restore Neurostimulator. The Restore Neurostimulator is an implanted, rechargeable

Recall record Z-0905-06 for Medtronic Inc, Neurological & Spinal Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0905-06
Company in FDA record
Medtronic Inc, Neurological & Spinal Division
FDA classification date
May 25, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A limited number of Model 37711 Restore Neurostimulators have an incorrect internal memory parameter that causes the device to indicate it is fully recharged when it may only be 80% to 90% recharged. This anomaly does not affect patient safety. This anomoly can be corrected by reprogramming the Restore Neurostimulator using specific software.

Source facts

Action
A 'Customer Communication' letter was sent to Medtronic field personnel to hand-carry into the affected accounts; customer contact began on 11/22/2005. The letter explains the patients implanted with these devices likely will only be able to recharge their neurostimulator to about 80-90% of a full charge. This anomaly does not affect device safety or effectiveness but Medtronic will re-program them upon request for customer satisfaction. The letter will be followed by scheduled on-site visit to reset the internal memory parameter of the neurostimulator.
Initiated
2005-11-22
Root cause
Other
Terminated
2006-12-26
Distribution
Nationwide (AZ, CA, FL, ID, IL, IN, KS, KY, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, and WI).
Recall number
Z-0905-06
Recall status
Terminated
CDRH record ID
43301
FDA classified
2006-05-25
Recall event number
34225

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗