Localization error; with planning data transferred to the Optical Guidance Platform via DICOM RT; The software incorrectly computes the center of the CT volume, resulting in a potential axial error ranging from 0.3 mm to 1.5 mm, affecting both Fractionated and Stereotactic Radiosurgery (SRS) Treatments
Source facts
Action
The recalling firm notified consignees by certified letter on 03/14/07. The notification was flagged as "Urgent Medical Device Correction" and advised end users that a localization error may occur when using the FramelessArray and affect its accuracy for Frameless treatments. Users were instructed to verify software in the devices and if running FramelessArray version 1.0, to discontinue use of the software when Stereotactic Radiosurgery plans are transfered via DICOM RT. The notification offered that firmware would be upgraded during scheduled service visit.
Initiated
2007-03-14
Root cause
Software design
Terminated
2008-07-03
Distribution
Worldwide : The software was distributed to direct accounts in the US and internationally distributed to accounts in Canada, Australia and Spain.
Recall number
Z-0905-2007
Recall status
Terminated
CDRH record ID
51590
FDA classified
2007-06-06
Recall event number
37811