Open original FDA record ↗Recall record
- FDA identifier
- Z-0905-2014
- Company in FDA record
- Medtronic Advanced Energy, LLC
- FDA classification date
- Jan 31, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use.
Source facts
- Action
- Medtronic sent an Urgent Product Removal Notification on November 12, 2013, via Next Day FEDEX. The communication advises users to immediately stop using the affected AQM3 electrosurgical generators and quarantine them until they can be returned to Medtronic Advanced Energy. Medtronic field personnel will collect them in order to remove them from service and return them to Medtronic Advanced Energy. Customers with questions were instructed to contact Customer Service at 866-777-9400. For questions regarding this recall call 866-777-9400.
- Initiated
- 2013-11-12
- Root cause
- Device Design
- Terminated
- 2014-04-03
- Distribution
- Nationwide Distribution including CA, MO, PA, NJ, NY, OH, VA, TX, NC, DE, KS, TN, LA, IL, MI, SC, WI, and FL.
- Recall number
- Z-0905-2014
- Recall status
- Terminated
- CDRH record ID
- 123689
- FDA classified
- 2014-01-31
- Recall event number
- 66835