Skip to content

Public FDA record

Z-0906-05: Ventak PRIZM 2 DR, model 1861 Automatic Implantable Cardioverter Defibrillator (AICD), Dual Chamber, Ste

Recall record Z-0906-05 for Guidant Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0906-05
Company in FDA record
Guidant Corporation
FDA classification date
Jun 30, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Laboratory analysis of returned devices revealed that deterioration in a wire insulator within the lead connector block, in conjuction with other factors, resulted in an electrical short. The short caused diversion of shock therapy energy away from the heart and into device circuitry.

Source facts

Action
Dear Doctor letter, dated 05/23/05, was sent to doctors 05/23/05 making them aware of 26 random component failures including one death. Doctors were advised to continue normal monitoring of all patients as indicated in device labeling. Press Release was also issued 05/25/05. Another Dear Doctor letter was dated and issued on 06/17/05. This letter provided updated information to the 05/23/05 letter. The letter gave some indications of device failure and recommendations for normal patient monitoring on routine follow-up visits. Another press release was also issued on 06/17/05.
Initiated
2005-06-17
Root cause
Other
Terminated
2007-01-05
Distribution
Worldwide. Including United States and countries such as: Arab Emirates, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Panama, Polynesia, Portugal, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Uruguay, Venezuela
Recall number
Z-0906-05
Recall status
Terminated
CDRH record ID
39930
FDA classified
2005-06-30
Recall event number
32379

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗