Open original FDA record ↗Recall record
- FDA identifier
- Z-0906-2007
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Jun 6, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Chair Breakage: Under extreme load, the chair back of the c.cam emission computed tomography system may bend and eventually break. This could be caused by the accumulated stress from heavy patients seating or stretching themselves in the chair or simply from long time use.
Source facts
- Action
- Important Safety Notice letters dated 4/30/07 on Siemens letterhead were sent to the end users, informing them of the potential risk that the c.cam chair may bend or break under heavy load. They were told that a Siemens service representative will contact them within the next few days to arrange the repair of the chair at no cost to the end user. To ensure the chair does not become damaged, SIemens recommended that the c.cam not be used for patient scanning until the update has been performed on the chair. Any questions regarding the Important Safety Notice were directed to their service representative or Siemens' office at: 1-800-888-7436 (USA).
- Initiated
- 2007-04-30
- Root cause
- Other
- Terminated
- 2009-01-09
- Distribution
- Worldwide-USA, Canada, France, Germany, Israel, The Netherlands and Portugal.
- Recall number
- Z-0906-2007
- Recall status
- Terminated
- CDRH record ID
- 51963
- FDA classified
- 2007-06-06
- Recall event number
- 37857