Skip to content

Public FDA record

Z-0906-2011: InfraVision IR Illuminator Console, model number 220-180-521, manufactured by Stryker Endoscopy San Jose,

Recall record Z-0906-2011 for Stryker Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0906-2011
Company in FDA record
Stryker Endoscopy
FDA classification date
Jan 18, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Product may not adequately visualize IR light emitted from U- and E- kits due to a manufacturing error.

Source facts

Action
The firm, Stryker Endoscopy, sent an "Urgent: Device Correction" letter dated September 28, 2010, t o all consignees/customers via a Fed Ex Mailer that included notification letter, acknowledgment receipts and updated instructions for use. International Stryker sites received product advisory notification. Stryker reps at those sites are responsible for notifying customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to discard any old IFUs from the InfraVision Console and ensure that they only use InfraVision compatible scopes; read the newly updated IFU provided with the letter; and complete and sign the enclosed "acknowledgement of receipt" form and fax to (408) 754-8378 or scan in or email to infravision@stryker.com. Note: Not returning the "acknowledgement of receipt" card will result in additional notices. Please contact your local Stryker sales representative to obtain information on scopes that are specific to the InfraVision system. If you have any questions, please contact us at 1-800-624-4422 or email at infravision@stryker.com.
Initiated
2010-09-29
Root cause
Nonconforming Material/Component
Terminated
2011-02-09
Distribution
Worldwide distribution: USA and countries including: Australia, Africa, Brazil, Canada, China, France, Greece, India, Italy, Korea, Latin America Mexico, New Zealand, Portugal, Romania, Singapore, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom.
Recall number
Z-0906-2011
Recall status
Terminated
CDRH record ID
94756
FDA classified
2011-01-18
Recall event number
56871

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗