Reason stated in the source

Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.

Source facts

Action
On 4/5/2023, correction notices were sent to customers who were asked to do the following: A permission-based feature called "Global Find" can be used to obtain alternate storage locations of medication(s) not loaded (by pharmacy) or depleted at that device and is not impacted by this issue. 1) Review and confirm that backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them to avoid delays in access to medications. 2) Ensure this letter is forwarded to applicable facilities if devices have been transferred and that all appropriate personnel have been made aware of this issue. 3) Complete and return the enclosed Customer Response Form via email to BDRC23@bd.com 4) Adverse health consequences experienced with the use of this product should be reported to BD at 1-800-727-6102 or via bd.com/self-service. Recall/Letter Questions: BD Customer Support, Phone: 1-866-583-8783, Phone Hours: 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET), Monday - Friday A software update is available. BD will contact you to schedule a time to install the software update.
Initiated
2023-04-05
Root cause
Software design
Distribution
US Nationwide distribution in the states of NY, NC, LA.
Recall number
Z-0906-2025
Recall status
Open, Classified
CDRH record ID
211732
FDA classified
2025-01-14
Recall event number
96002

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗